    Remove the black protection and turn the meter on with the ON / OFF button located above.
    Plunge it to the limit and wait for the data to stabilize.
    The meter automatically compensates for temperature variations.
    After use, run the electrode through water to keep it clean, and dry it
    The data communicated by the screen must be multiplied by 100 to give a direct reading in mS / cm
    Calibration:
    The calibration procedure is simple and fast
    Immerse the meter to the limit in EC calibration liquid (M10031)
    Wait for the data to stabilize and with the screwdriver to the screw on the back,
    Set to 1.4 (multiplied by 100 gives an EC of 1413). The meter is calibrated.
    ALWAYS USE new calibration fluids.
    Specs:
    Range? S 0-1990 / cm
    Resolution 1 S / cm
    Accuracy ▒ 2% of its capacity
    Manual Calibration 1 point
    Temp. Automatic compensation of 5 - 50 ░ C
    Battery life 350 hours
    Dimensions 150 x 30 x 24 mm