The MW101 PH meter is a portable PH meter with 1 meter cable probe, BNC connector. Measurement range is 0.01 to 14.01 pH; Accuracy 0.01pH
Manual calibration, in 2 points (PH7.01 and PH 4.01 / 10.01); Manual temperature compensation. Wide screen, ergonomic shape, safety strap; The electronics of the MW101 PH meter are modern for fast and efficient measurement.
The PH probe (ref: SE220) is a double junction electrode filled with electrolyte solution. Electrode suitable for MW100 - MW101 - MW102.
The MW101 PH meter is ideal for field measurements. It adapts very well to the sectors of agriculture, hydroponics, aquariums, fish farming, education, laboratories, swimming pools.
MW101 Precision PH Meter is a portable PH meter with 1 meter cable probe, BNC connector.
The measurement range of the MW101 precision PH meter is from 0.01 to 14.01 pH; Accuracy 0.01pH
The precision pH meter MW101 is calibrated manually, mind in 2 points (PH7.01 and PH 4.01 / 10.01); Manual temperature compensation. Wide screen, ergonomic shape, safety strap; The electronics of the MW101 PH meter are modern for fast and efficient measurement.
The PH probe (ref: SE220) is a double junction electrode filled with electrolyte solution. Electrode suitable for MW100 - MW101 - MW102.
The MW101 precision PH meter is ideal for field measurements.
The MW101 precision PH meter is very well suited to agriculture, hydroponics, aquariums, fish farming, education, laboratories, swimming pools sectors.
Operation of the MW101 precision PH meter
The MW101 precision PH meter is delivered complete, with a 9V battery. Open the battery compartment cover on the back. Uncertain the battery respecting the good polarity.
Always remove the electrode protection before use. In the event that the electrode was left dry, immerse it (up to 2.5cm) for a few minutes in washing liquid to reactivate it.
Connect the electrode to the socket located above the meter.
Make sure the meter was calibrated before use.
Immerse the pH electrode (2.5 cm) in the provided liquid.
Turn on the meter by pressing the ON / OFF button
Wait for the data to stabilize before taking the data for granted.
After the measurement, turn off the meter and replace the cap to the electrode with a few drops of storage liquid inside.
MW101 Precision PH Meter Calibration:
Remove the protection from the electrode.
Immerse the electrode in an envelope of pH7.01 calibration liquid and wait for the data to stabilize.
Using the black PH7 calibration button, adjust the data to pH7.0
Open an envelope of pH4.01 calibration liquid and use a small amount of it to wash the electrode.
Immerse the electrode in the pH4.01 liquid envelope and read the data once stabilized.
Set the PH4 / PH10 button to PH4.0 data.
The meter is calibrated and ready to measure.
It is advisable to calibrate the meter at least once a month, or after a long period of non-use or after changing the electrode.
For any clarification on the MW101 precision PH meter, do not hesitate to contact us. Good measurements!