    Milwaukee Waterproof Ph Meter (PH51) Manual Calibration

    PHEC0133

    Sold by Natural Systems. More information

    The Milwaukee Waterproof pH Meter (PH51) is a device used to measure the pH of water accurately and reliable. It is waterproof, making it suitable for use in aquatic environments or in harsh conditions. wet.
    One of the important features of this meter is that it allows manual calibration at two points: acid and alkaline. Manual calibration is a necessary process to ensure the accuracy of pH measurements.

    Conditioning:

    Remove the cap and immerse the pH51 in MA9015 maintenance solution for 2 hours to activate the electrode

    CALIBRATION:

    Unscrew the battery compartment located on the top of the pH-meter and perform the following two-point calibration procedure:

    Acid calibration: for pH measurements below 7 pH:

    • Immerse the electrode in the buffer solution pH7.01 (M10007)
    • Allow the reading to stabilize and then, using the screwdriver, turn the calibration screw "7" until the reading reads "7.0".
    • Rinse the electrode with tap water. Immerse the electrode in the buffer solution pH4.01 (M10004).
    • Allow the reading to stabilize and then, using the screwdriver, turn the calibration screw "4" until the reading reads "4.0".

    Alkaline calibration: for measurements above pH7:

    • Immerse the electrode in the buffer solution pH7.01 (M10007)
    • Allow the reading to stabilize and then, using the screwdriver, turn the calibration screw "7" until the reading reads "7.0".
    • Rinse the electrode with tap water. Immerse the electrode in the buffer solution pH10.01 (M10010).
    • Allow the reading to stabilize and then, using the screwdriver, turn the "10" calibration screw until the readout reads "10.0".
    Technical specifications:
    • Water resistant
    • Range 0.0 - 14.0 pH
    • Resolution 0.1pH
    • Accuracy ±0.1pH
    • Manual calibration 2 points
    • MA73600 Electrode (Replaceable Probe)
    • Battery 3 x 1.5V
    • Duration 1500 hours
    • Storage 0 to 50 °C/95% RH max

    Changing batteries: The pH51 has the BEPS system (battery error prevention system) that prevents the reader from taking wrong measurements due to dead batteries, turning off the instrument:

    • To change the batteries, unscrew the battery compartment and replace the three 1.5 V batteries, paying attention to the polarity of these.
    • Make sure the ring is in place before screwing on the compartment.
    • Batteries should only be changed in an area not dangerous and with the type of battery specified in this manual.